Crucible furnace part | |
Model | PT-V1400-6LZ |
crucible size | diameter 200*200mm |
Crucible Material | Corundum Crucible |
Working temperature | ≦1300℃ |
Maximum.temperature | 1400℃ |
Temperature Control System | 7-inch touch screen intelligent operation, can store 16 process curves, each process can be programmed with 30 segments, and can record temperature data in real time, which can be copied to the computer |
Constant temperature accuracy | ±1℃(high temperature heat preservation stage) |
Over temperature protection | With over temperature and broken couple alarm function |
heating rate | 0-20℃/min |
Heating element | High temperature silicon carbide rod |
Temperature measuring element | S-type thermocouple temperature measurement |
Furnace material | Using polycrystalline fiber, the material has good thermal insulation performance, high reflectivity and balanced temperature field |
shell structure | Double-layer shell, inner cylindrical structure pressure-resistant, outer square shell |
top seal | Stainless steel sealing flange, sealed with sealing ring, and the air inlet is connected to the air circuit. |
Customized part | A gas collection chamber is added at the gas outlet end to collect gas, and the size is φ100*200mm |
vacuum system part | |
Name | Rotary Vane Pump |
The main parameters | Pumping speed 6.6L/S Vacuum degree 10Pa |
vacuum display | Inficon Imported digital vacuum gauge, anti-corrosion, accurate reading |
capture system | φ100mm×200mm Stainless steel tube, material holder device |
filter device | The pump port is equipped with a filter device to filter the dust and oil pollution, etc., to avoid contamination of the pump oil. |
Air system part | |
Name | Two-way proton flowmeter |
Gas line | Two-way |
flow control | 7-inch LCD touch screen control, digital display, each gas containing needle valve is individually controlled |
link method | Double ferrule joint |
Flowmeter | Mass flowmeter |
Flow range | Nitrogen: 0-1000 sccm Hydrogen: 0-100 sccm |
mixed gas | Contains a precision mixing tank |
Working temperature | 5℃-45℃ |