Product name | Tube PECVD system (CE certification) |
Slide tube furnace part | |
Model | PT-T1200 |
Heating zone length | 300mm |
Furnace tube size | diameter 120*1600mm |
Working temperature | ≦1100℃ |
Max. temperature | 1200℃ |
Temperature measuring element | K type thermocouple |
Temperature Control System | 30-segment PID microcomputer programmable automatic control |
Insulation temperature control accuracy | ±1℃ |
temperature protection | With over temperature and burnout protection function |
Heating rate | 0-20ºC/min |
Heating element | High resistance electric heating alloy wire OCr27A17Mo |
Operating Voltage | AC220V single phase 50HZ |
rated power | 4KW |
chamber material | Using polycrystalline fiber furnace material, the material has good thermal insulation performance, high reflectivity and balanced temperature field |
sealing system | The furnace tube and the flange are squeezed and sealed with O-rings, which are easy to disassemble, can be disassembled repeatedly, and have good air tightness. |
Furnace tube | High purity quartz tube |
French | 304 stainless steel sealing flange, installed mechanical pressure gauge and stainless steel globe valve, easy to disassemble and take out materials |
Furnace structure | Cr-plated thick steel plate two-way slide rails can realize the movement of the tube furnace to achieve rapid heating and cooling |
Furnace shell temperature | ≤45℃ |
vacuum system part | |
name | DRV16 Rotary Vane Pump 4L/S |
The main parameters | Ultimate vacuum degree 10Pa (including KF25 interface, vacuum baffle valve and movable box) |
vacuum display | Resistance digital display vacuum gauge, display range standard atmospheric pressure--10-1pa |
Air system part | |
name | Four-way mass flow meter |
Gas line | four way |
flow control | Digital display, each channel of gas contains a needle valve for separate control, and the gas flow is automatically controlled |
Air inlet and outlet | Four in and one out, built-in precision mixing tank |
link method | Double Ferrule Fitting |
flow range | MFC1-MFC4:0-500sccm(range optional) |
control precision | About ±1.5%F.S, different ranges, slightly different |
Working temperature | 5℃-45℃ |
Plasma radio frequency part | |
name | RF automatic matcher |
Matching power range | 0-500W |
working frequency | 13.56MHZ+0.005% |
Normal working reflected power | <3W |
match time | <3S |
transmission efficiency | >90% |
Matching power supply | AC220V/50-60HZ |
We provide 24/7 hours global online service, one-stop customized service, welcome to consult.