Can put inner tube into the bid size outer tube, make big-area monolayer graphene growing test.
The device widely used in the field of semi-conductor, nanometer material, carbon fibre, graphene, etc. with cost-effective feature.Working Temperature | Max. Temperature: 1000℃ ( 1 hr) Continuous working temperature: ≤1000℃ Recommended heating Rate: ≤10℃/min |
Temperature Controller | PID automatic control and auto-tune function. 30 programmable segments for precise thermal processing. Built-in protection for the over-heated and broken thermocouple. |
Length of heating zone | 300+300+300mm |
Length of constant zone | 150+200+150mm |
Heating Elements | Resistance wire, Fe-Cr-Al Alloy doped by Mo |
Temperature Accuracy | ±1℃ |
Dimensions | Tube size: Φ150/210/280*1400mm (optional) Overall size: 1300*650*1000mm Two fibrous ceramic tube blocks are included for blocking heat radiation generated from tube center. (Ceramic blocks must be fully inserted into the furnace chamber before heating).Quartz blocks also optional |
Vacuum Flange |
Folding flanges |
Vacuum Pressure | 133torr can be achieved by mechanical pump 10^-6 torr can be reached by molecular pump Please click the picture below to order vacuum pump |
Warranty | One year limited warranty with lifetime support (Such consumable parts as tube, quartz block, and O-rings, etc. are NOT supported by the warranty) |
Warning | Small cracks may appear on the surface of the refractory ceramics over extended use. This is a normal occurrence and the cracks may be repaired with alumina coating. |